User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Keithley 197: (90 day) CAL VER DATE: 30-Jul-92 AUTHOR: User Contributed REVISION: 2.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 68 NUMBER OF LINES: 131 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F W 1.002 ASK+ B 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Allow at least 1 hour warm up before testing. 1.005 DISP [32] Connect the 5700A and the UUT as follows. 1.005 DISP [32][27][91]1m 5700A TO 197 1.005 DISP [32] OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ INPUT HI 1.005 DISP [32] OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ INPUT LO 1.005 DISP [32] SENSE V ê HI ÄÄÄÄÄÄÄÄÄÄÄ ê SENSE HI 1.005 DISP [32] SENSE V ê HI ÄÄÄÄÄÄÄÄÄÄÄ ê SENSE LO 1.006 DISP [27][91]1m[27][91]5m!NOTE! 1.006 DISP If the UUT fails any portion of the following tests, 1.006 DISP select ([27][91]1mK[27][91]mBRD) or (K) instead of (NO). 1.006 DISP This will enable data of the failure to be recorded. 1.007 HEAD {DC VOLTS TESTS} 1.008 RNG 200mV 1.009 MESS On UUT press the [27][91]1mREL[27][91]m pushbutton 1.009 MESS and verify the display indicates "REL". 1.010 5700 0V 0.001U 2W 2.001 MESS 2.002 5700 200 200mV 0.035U 2W 3.001 5700 200 -200mV 0.035U 2W 4.001 5700 2 2V 0.00024U 2W 5.001 5700 2 -2V 0.00024U 2W 6.001 5700 20 20V 0.0032U 2W 7.001 5700 20 -20V 0.0032U 2W 8.001 5700 200 200V 0.032U 2W 9.001 5700 200 -200V 0.032U 2W 10.001 5700 1000 1000V 0.32U 2W 11.001 5700 1000 -1000V 0.32U 2W 12.001 HEAD {AC VOLTS TESTS} 12.002 DISP Press [27][91]1mREL[27][91]m pushbutton and verify 12.002 DISP display does not show "REL". 12.003 5700 200 200mV 2.1U 20H 2W 13.001 5700 200 200mV 0.8U 50H 2W 14.001 5700 200 200mV 1.4U 10kH 2W 15.001 5700 200 200mV 3.25U 20kH 2W 16.001 5700 200 200mV 10.4U 50kH 2W 17.001 5700 200 200mV 10.4U 100kH 2W 18.001 5700 2 2V 0.021U 20H 2W 19.001 5700 2 2V 0.008U 50H 2W 20.001 5700 2 2V 0.014U 10kH 2W 21.001 5700 2 2V 0.0325U 20kH 2W 22.001 5700 2 2V 0.064U 50kH 2W 23.001 5700 2 2V 0.064U 100kH 2W 24.001 5700 20 20V 0.21U 20H 2W 25.001 5700 20 20V 0.08U 50H 2W 26.001 5700 20 20V 0.14U 10kH 2W 27.001 5700 20 20V 0.325U 20kH 2W 28.001 5700 20 20V 0.64U 50kH 2W 29.001 5700 20 20V 0.64U 100kH 2W 30.001 5700 200 200V 2.1U 20H 2W 31.001 5700 200 200V 0.8U 50H 2W 32.001 5700 200 200V 1.4U 10kH 2W 33.001 5700 200 200V 3.25U 20kH 2W 34.001 5700 200 200V 6.4U 50kH 2W 35.001 5700 1000 750V 5.62U 50H 2W 36.001 5700 1000 750V 9.5U 10kH B1 2W 37.001 HEAD {DC CURRENT TESTS} 37.002 5700 200 200uA 0.215U 2W 38.001 5700 2 2mA 0.00215U 2W 39.001 5700 20 20mA 0.0215U 2W 40.001 5700 200 200mA 0.415U 2W 41.001 5700 2000 2000mA 4.15U 2W 42.001 DISP Remove all connections. 42.002 DISP [32]Connect the 5725A and UUT as follows: 42.002 DISP [32][27][91]1m 5725A TO 197 42.002 DISP [32] CURRENT OUTPUT HI ÄÄÄÄÄÄÄÄÄÄÄÄ INPUT 10A 42.002 DISP [32] CURRENT OUTPUT LO ÄÄÄÄÄÄÄÄÄÄÄÄ INPUT LO 42.003 5700 10 5A 0.039U B1 2W 43.001 HEAD {AC CURRENT TESTS} 43.002 5700 10 5A 0.085U 40H B1 2W 44.001 5700 10 5A 0.06U 400H B1 2W 45.001 5700 10 5A 0.06U 1kH B1 2W 46.001 DISP Remove all connections. 46.002 DISP [32] Connect the 5700A and the UUT as follows. 46.002 DISP [32][27][91]1m 5700A TO 197 46.002 DISP [32] OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ INPUT HI 46.002 DISP [32] OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ INPUT LO 46.003 5700 200 200uA 2.1U 20H 2W 47.001 5700 200 200uA 1.7U 50H 2W 48.001 5700 200 200uA 4.25U 10kH 2W 49.001 5700 2 2mA 0.021U 20H 2W 50.001 5700 2 2mA 0.017U 50H 2W 51.001 5700 2 2mA 0.0425U 10kH 2W 52.001 5700 20 20mA 0.21U 20H 2W 53.001 5700 20 20mA 0.17U 50H 2W 54.001 5700 20 20mA 0.425U 10kH 2W 55.001 5700 200 200mA 2.1U 20H 2W 56.001 5700 200 200mA 1.7U 50H 2W 57.001 5700 200 200mA 1.73U 10kH 2W 58.001 5700 2000 2000mA 21U 40H 2W 59.001 5700 2000 2000mA 17U 400H 2W 60.001 5700 2000 2000mA 17.3U 10kH 2W #! WARNING: Test Tol 0.0173, Sys Tol 0.0182, T.U.R. 0.951 (< 4.00). 61.001 HEAD {4-WIRE RESISTANCE TESTS} 61.002 DISP When using ê function, UUT AC/DC switch must be in 61.002 DISP DC position or "Err" will be displayed. 61.003 DISP [32] Connect the 5700A and the UUT as follows. 61.003 DISP [32][27][91]1m 5700A TO 197 61.003 DISP [32] OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ INPUT HI 61.003 DISP [32] OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ INPUT LO 61.003 DISP [32] SENSE V ê HI ÄÄÄÄÄÄÄÄÄÄÄ ê SENSE HI 61.003 DISP [32] SENSE V ê HI ÄÄÄÄÄÄÄÄÄÄÄ ê SENSE LO 61.004 5700 200 190Z 0.041U 4W 62.001 5700 2 1.9kZ 0.00036U 4W 63.001 5700 20 19kZ 0.0051U 4W 64.001 5700 200 190kZ 0.051U 4W 65.001 ASK- R 65.002 DISP Set UUT to the Mê range. 65.003 5700 1.9MZ 0.00069U 4W 66.001 5700 19MZ 0.023U 4W 67.001 5700 100MZ 2.001U 2W 68.001 DISP {C A L I B R A T I O N C O M P L E T E !} 68.002 HEAD 68.003 END #! T.U.R.s less than 4.00: 1 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.